Phytophthora ramorum is the causal agent of sudden oak death in forests on the west coast of the United States, and dieback and leaf blight in a wide range of woodland trees, shrubs, and  herbaceous plant species in the UK and other areas of Europe. Suitable detection methods are  needed in order to facilitate effective measures for the control and eradication of this pathogen. In particular, rapid methods that can be carried out in the field, minimizing the delay between sampling and diagnosis, could help to reduce the spread of P. ramorum. Lateral flow devices can be used to identify members of the genus Phytophthora in the field, but identification at the species level requires a nucleic acid-based test.

 

 

Figure 1. Phytophthora ramorum on Viburnum sp. Leaf

 

 

 

 

A loop-mediated isothermal amplification (LAMP) assay using six primers (annealing to 8 regions) was developed for the detection of P. ramorum in plant material. The sensitivity and specificity of this method are similar to those of real-time PCR methods for detection of this pathogen, with the advantages that LAMP does not require thermal cycling and the reaction is completed in 1 hour. LAMP assays have subsequently been developed at Fera for other plant pathogens, including P. kernoviae and Botrytis cinerea.

LAMP products can be detected in the laboratory by gel electrophoresis, or in the field by observation of precipitated magnesium pyrophosphate or a colour change on the addition of an intercalating dye. However, these methods require the reaction tubes to be opened after amplification, which can result in carry-over contamination. Given that a LAMP reaction generates of the order of 25 μg of amplicon, five to ten more than a standard PCR, it is very important to avoid opening the finished product reaction tube.

Figure 2. Sensitivity and specificity of Phytophthora ramorum LAMP assay. A. A dilution series of P. ramorum DNA amplified by LAMP. M: marker; lane 1: negative control (water); lanes 2 and 3: 10 ng; lanes 4 and 5: 1 ng; lanes 6 and 7: 100 pg; lanes 8 and 9: 50 pg; lanes 10 and 11: 10 pg; lanes 12 and 13: 1 pg.

B. Cross reactivity of LAMP assay with P. lateralis DNA. M: marker; lane 1: negative control (water); lane 2: positive control (10 ng P. ramorum DNA); lanes 3 and 4: 70 ng P. lateralis DNA; lanes 5 and 6: 7 ng P. lateralis DNA; lanes 7 and 8: 700 pg P. lateralis DNA

The GENIE detection system developed and manufactured by OptiGene allows real-time LAMP to be performed on a low-power portable platform, suitable for use in the field. The closed-tube nature of GENIE avoids any post-amplification handling thereby eliminating laboratory contamination with the amplified product.

The GENIE system consists of two independent heating blocks, each block taking up to eight 0.2ml microtubes and offering a gradient facility for optimisation of the LAMP reaction temperature.

 

Figure 3. Time to positive result for a dilution series of P. ramorum DNA on the GENIE detection system

 

 

Figure 5. Amplification plots for real-time LAMP performed on the GENIE detection System

 

GENIE measures the fluorescence, in real time, of an intercalating dye during the amplification process. This allows the reaction to be monitored and the time of initial amplification gives an indication of template copy number.

From melt curves LAMP reactions primarily consist of a single amplified product usually melting between 83 – 89°C. This is several degrees higher than one sees for standard PCR products. It has been found that by performing an annealing curve on the finished amplification product one sees a single product peak from which one can extract a Ta value. Similar to a melt curve, this annealing Ta value is unique to the sequence amplified, thus confirming amplification of the correct product.

 

Figure 5. Melt curve analysis on the GENIE detection system

The use of fast isothermal amplification reagents using a novel Geobacillus sp. Large Fragment DNA polymerase allows results to be obtained in less than 30 minutes.

DNA extracted from a culture of P. ramorum tested positive by real-time LAMP in as little as 12 minutes in contrast with LAMP using NEB Bst Large Fragment DNA polymerase which typically takes around 1 hour.

References: Tomlinson J.A., Barker I., and Boonham N. 2007. Faster, simpler, more-specific methods for improved molecular detection

of Phytophthora ramorum in the field. Applied and Environmental Microbiology 73:4040-4047.

Newsletter Signup
Be The first to hear the latest news from Optigene

twitter-button

    Free Sample Request Form

    One sample per organisation - new customers only.
    Isothermal Master Mix samples will be sent as dried reagents only. These require re-suspension before use. Dried reagents perform near identically to the wet reagents but do not require cold shipment or storage.
    Both wet and dry reagents are available for purchase from OptiGene.

     
    Please only press 'Submit' once.
    You will receive a sample as soon as possible.